Disposable GIA Linear Cutter Stapler
I. Solution Description
The disposable linear cutter is applicable for transection, resection, and the generation of anastomoses. It has 4 rows of staples with a knife in between the 2nd and third row. It is a sterile, single-individual-use instrument that simultaneously staples and divides tissue
The YQG series linear cutter is a single-patient-use device. It provides 4 staggered rows of titanium staples, with a knife in between the next and 3rd row. It can be operated by 1 hand. Two-way press knob can be suited for remaining or appropriate hand.
We have two kind of staple peak cartridges which supply the flexibility to produce acceptable compression for the targeted tissue.
II. Scope of Software
It is applicable to mutilation, excision and suture in the procedure of alimentary canal such as belly and intestines.
III. Exterior Construction
IV. Contraindication
1. Serious mucous edema.
two. Never use the stapler with staple height of 4.5mm for any tissue whose compressed thickness is considerably less than 2.0mm, due to the fact the stapler staple are not able to completely tighten up to stop bleeding in this circumstance.
3. Never use the stapler with staple top of four.5mm for any tissue which fails to be compressed effortlessly to be 2.0mm thick, since the tissue thickness is beyond the suturing specification in this scenario.
4. Never ever use the stapler with staple peak of 3.85mm for any tissue whose compressed thickness is considerably less than one.5mm, due to the fact the stapler staple cannot entirely tighten up to stop bleeding in this circumstance.
five. By no means use the stapler with staple peak of three.85mm for any tissue which fails to be compressed very easily to be one.5mm thick, simply because the tissue thickness is outside of the suturing specification in this circumstance.
six. In no way use the instrument for any tissue whose compressed thickness is much less than 1mm, since the stapler staple are not able to totally tighten up to cease bleeding in this case.
7. The instrument shall not be employed for livers or spleens, simply because closing the instrument may lead to destructive result because of to their compression qualities.
V. Instructions for Operation
1. Consider the instrument out of the package deal in the sterile purifying location. To steer clear of any harm, the instrument shall be handled gently.
two. Open the free of charge manage go over, different the cutter assembly from the anvil, and then eliminate the cartridge protect.
three. Clamp the tissue to be minimize and sutured in between the cartridge and the anvil assist, and then press the free deal with cover to the locking situation (Observe: The free of charge deal with cover should be pressed to its spot).
four. Maintain the cutter stapler in 1 hand and press the ejector pad with the thumb and forefinger of the other hand to the firing placement (as the reducing knife nonetheless does not start reducing, the hand really feel is gentle), and then proceed pushing the ejector pad uniformly forwards to the front end (the hand truly feel is hefty, indicating that the reducing knife starts off reducing the tissue and suturing the tissue at equally sides).
5. Reset the ejector pad (there is pink sign on the instrument, as revealed in figure 1). Pay out consideration that the stapler are not able to be lifted quickly. The free handle include have to be first unveiled and the cartridge and anvil support shall be separated right after that, the instrument shall be taken out, and then the operation is completed.
Removable assembly:
Use of removable assembly of disposable linear cutter stapler and cartridge
Be positive that the ejector pad returns to its authentic position, and then pull the totally free take care of include open to remove the utilised assembly and exchange it with a new one place the new assembly in the assembly holder and repair it, and then take away the include the assembly can be utilised in the exact same procedure according to the utilization.
VI. Storage
The item shall be stored in a well-ventilated room where no corrosive gases exist and relative humidity is not a lot more than eighty%.
VII. Company and Its Handle and Speak to Amount
Company: HangJCPT JCPT JCPT Deveices
Deal with: No. 2188 Xihu (West Lake) Dis. Rd, Xihu (West Lake) Dis. District, HangJCPT City China
Zip Code: 213012
Design | Cartridge Product | Chopping Size | Staple Height(mm) | Shade of Cartridge | Suture Size(mm) | Suture Thickness(mm) |
YQG-55L | YQGZ-55L | 55 | 3.85 | Blue | 60 | 1.5 |
YQG-55H | YQGZ-55H | 55 | 4.fifty | Green | 60 | 2. |
YQG-75L | YQGZ-75L | seventy five | three.85 | Blue | 80 | 1.5 |
YQG-75H | YQGZ-75H | seventy five | 4.fifty | Eco-friendly | eighty | two. |
YQG-100L | YQGZ-100L | one hundred | 3.85 | Blue | one hundred and five | 1.5 |
YQG-100H | YQGZ-100H | 100 | four.50 | Inexperienced | one hundred and five | 2. |
VIII. Bundle
IX. Composition Figures
1. Staple Row Extends About 1.five Staples Outside of Minimize Line.
2. Reloading cartridge
Blue Cartridge: one.5mm tissue
Environmentally friendly Cartridge: 2.0mm tissue
three. Tissue-retaining Pin guarantee accurate reducing and suturing
four. Intermediate locking position allows maneuvering and repositioning of tissue inside instrument just before fining.